Page 46: Automatic, Automatic switch-off function, Display illumination, Section 7.3.3 a, Utomatic, Tability, Trol

Advertising
background image

Settings

HandyLab 600

46

ba77064e01

03/2014

7.3.3 Automatic Stability control

The automatic Stability control (AutoRead) function continuously checks the
stability of the measurement signal. The stability has a considerable impact on
the reproducibility of measured values.

You can activate or switch off the automatic Stability control function (see
section 7.3 S

ENSOR

-

INDEPENDENT

SETTINGS

, page 44).

The measured parameter flashes on the display

as soon as the measured value is outside the stability range

when the automatic Stability control is switched off.

7.3.4 Automatic switch-off function

The instrument has an automatic switch-off function in order to save the
batteries (see section 7.3.1 S

YSTEM

, page 44). The automatic switchoff function

switches off the meter if no key is pressed for an adjustable period.

The automatic switchoff function is not active

if a USB-B cable is connected

if the Automatic data storage function is active, or with automatic data trans-
mission

7.3.5 Display illumination

The meter automatically switches off the display illumination if no key is
pressed for 20 seconds.

The illumination is switched on with the next keystroke again.

You can also generally switch on the display illumination (see section 7.3.1
S

YSTEM

, page 44).