Meter does not react
to keystroke
You want to know
which software
version is in the
meter or IDS sensor
Cause
Remedy
– Operating condition undefined
or EMC load unallowed
– Processor reset:
Press the <ENTER> and <On/Off>
key simultaneously
Cause
Remedy
– E.g., a question by the service
department
– Switch on the meter.
– Open the menu, <ENTER_> / Stor-
age & config / System / Service infor-
mation. The instrument data are
displayed.
or
– Connect the sensor.
Press softkey [<F1>Info<F1>] /
[More] The sensor data are displayed
(see section 4.1.5 S