Xylem HandyLab 600 User Manual

Page 32

Advertising
background image

pH value

HandyLab 600

32

ba77064e01

03/2014

Calibration record

(USB output)

-

-30 ... <-25
or
->+25 ... +30

-62.0 ... <-61.0
or
>-56.0 ... -50.0

Clean the IDS sensor according to the sensor operating manual

Error

Error

<-30
or
>+30

<-62.0
or
> -50,0

Error elimination (see section 14 W

HAT

TO

DO

IF

..., page 88)

For pH IDS sensors you can optionally enable a more finely graded
calibration evaluation (QSC) (see section 5.2.8 QSC

FUNCTION

(

SENSOR

QUALITY

CONTROL

), page 34).

Display

Calibration
record

Zero point
[mV]

Slope [mV/pH]

CALIBRATIONpH
01.02.2014 07:43:33

BlueLine 24 pH IDS
Ser. no. B092500013

Buffer 1
Buffer 2

7.00

Buffer 3
Voltage

1

184.0 mV 24.0 °C

Voltage 2

3.0 mV 24.0 °C

Voltage

3

-177.0 mV 24.0 °C

Slope

-60.2 mV/pH

Asymmetry

4.0 mV

Sensor

+++

etc...