Xylem HandyLab 600 User Manual

Page 26

Advertising
background image

pH value

HandyLab 600

26

ba77064e01

03/2014

Continuing with

three- to five-point

calibration

12. Start the measurement with <ENTER>.

The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.

13. Wait for the measurement with stability control to be completed or ter-

minate the stability control and take over the calibration value with
<ENTER>.
The calibration display for the next buffer appears (voltage display).

14. If necessary, finish the calibration procedure as a two-point calibration

with <M>.
The calibration record is displayed.

pH

0.0

24.8 °C

7.000

01.07.2011

08:00

AR

Buffer

mV

15. Thoroughly rinse the sensor with deionized water.

16. Immerse the sensor in the next buffer solution.

17. Start the measurement with <ENTER>.

The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.

18. Wait for the measurement with stability control to be completed or ter-

minate the stability control and take over the calibration value with
<ENTER>.
The calibration display for the next buffer appears (voltage display).

pH

180.0

24.8 °C

01.07.2011

08:00

AR

Buffer

mV