3 sensor monitoring, Senso, E section 6.4.3 – YSI DIQ/S 182 XT-4 System 182 XT-4 User Manual

Page 109

Advertising
background image

System 182 XT-4

Outputs

6 - 17

ba76026e01

01/2012

6.4.3

Sensor monitoring

Function

The Sensor monitoring function enables the monitoring of sensor errors
and the maintenance condition.

In order to set up the Sensor monitoring function for a relay output, the
relay output must be linked with a sensor (see section 6.2.1).

Settings

Note
Preferably set up the relay output as an opener for monitoring
functions.

Sensor messages include errors and information that are registered by
the sensor.

Special

sensor errors

The special sensor errors are sensor-dependent. Details of this are
given in the component operating manual of the respective sensor.

General

sensor errors

Setting

Selection

Explanation

Errors

All

All sensor errors (special ones
and general ones) are
monitored and can prompt a
relay action.

Special

Special sensor errors are
monitored and can prompt a
relay action.

Off

Sensor errors are not
monitored.

Mainten. condition

On
Off

Switching on and off the
maintenance condition is
monitored and can prompt a
relay action.

Action

Open

For all functions of the Sensor
monitoring
, the relay action is
set to Open.

Init

This can prompt a relay action for a short time,
depending on the starting behavior of the system

----

Invalid measured value, or defective sensor

Error

Communication with sensor interrupted

OFL

Measuring range undercut or exceeded (overflow)