Page 10
Advertising
Overview
System 182 XT-4
1 - 4
ba76026e01
01/2012
1.2
Behaviour of the system to a power failure
The system configuration is non volatile and is permanently
retained. It consists of the following settings:
– Sensor settings
– Settings and links of the relay outputs
– Settings and links of the current outputs
– System settings (display language, air pressure/altitude,
passwords etc.)
Linked relay outputs switch to the non active condition (open).
Linked current outputs switch to the non active condition (0 mA).
A restart is carried out automatically when the power returns. The
system continues to work with the settings at the time of the power
failure (except for the time).