System 182 XT-4

Outputs

6 - 31

ba76026e01

01/2012

Within the proportional range an increase of the concentration by 1 mg/
l causes a reduction of the output current by 1 mA. If the measured
concentration is 17 mg/l, for example, 11 mA is output:

The highest concentration at which the controller still works in the
proportional range is 8 mg/l (corresponding to Imin = 8 mA) and the
lowest is 2 mg/l (corresponding to Imax = 14 mA).

Settings

(

)

mg/l

mg/l

mg/l

mA

1

mA

12

I

Controller

5

4

+

=

(

)

mA

11

mg/l

1

mg/l

mA

1

mA

12

I

Controller

=

+

=

Setting

Selection/Values

Explanation

Measured variable

Main variable
Adjoining variable

Main variable
designates the actual
measured parameter
of the sensor (e.g.
pH, oxygen, etc.).
Adjoining variable
designates an
additional measured
parameter (e.g.
temperature).

Nominal value

Within the measuring
range (sensor-
dependent)

Nominal value the
measured value is
regulated to

Xp

5 ... 100 %
-5 ... -100 %
% of the measuring
range

Proportional range of
the controller.
Negative values
result in a positive
characteristic curve.

Imin

0 ... 20 mA

Lower current
limitation *

Imax

0 ... 20 mA

Upper current
limitation *

* Note:
Spacing between
Imin and Imax:
at least 5 mA