Technical data

System 182 XT-4

9 - 4

ba76026e01

01/2012

Relay

(6 x)

Current outputs

(5 x)

Output

Galvanically separated

Max. switching voltage

Front terminal strip:

240 VAC or 24 VDC

Rear terminal strip:

250 VAC or 24 VDC

Max. switching current

Front terminal strip:

2 A (AC and DC)

Rear terminal strip:

5 A (AC and DC)

Fuse rating on the
operator side

Front terminal strip:

max. 2 A

Rear terminal strip:

Maximum 5 A

Relay functions

Programmable as:

– Opener or closer

– Limit monitor

– Monitoring of the warning and error

signals of the DIQ/S 182 XT-4

– Proportional frequency output

– Proportional pulse width output

Output

Galvanically separated from the sensors

Output current

Can be switched between 0 - 20 mA and
4 - 20 mA
In the case of errors, can be set to: 0 ... 21
mA

Max. initial output
voltage

24 V, in the case of missing or incorrect
burden

Accuracy

0.3 % of current value ± 50 µA,
load max. 500

Ω

Functions

Programmable:

– Recorder attenuation, adjustable

0 - 40 mA/s

– Error behavior can be set as required

0 ... 21 mA

– Error behavior according to Namur NE43

can be selected

– Positive and negative characteristic

curve

– PID controller