What to do if ...

System 182 XT-4

8 - 4

ba76026e01

01/2012

Note
It is always possible to replace components and assign a substitute if
the software state of the substitute component is as high or higher as
the software version of the original component. Otherwise the
component may not be accepted as a substitute component by the
DIQ/S 182 XT-4.

Operating sequence in

case 2

1

Connect the IQ sensor.

2

Change to the measured value display with

m

. The module

database is updated. The following display appears (example):

3

Select the required option with

w

and confirm with

g

.

– If Add as new sensor was selected, the system changes

directly to the measured value display. As soon as the
IQ sensor is ready for operation, it delivers a measured
value.

– If Substitute sensor was selected, a list with the assignable

sensors or inactive datasets appears.

4

Select the required sensor with

w

and confirm with

g

. The

system changes to the measured value display. The IQ sensor
takes over all settings of the inactive dataset. As soon as the
IQ sensor is ready for operation, it delivers a measured value.