4 distributed mounting, Distributed mounting -18 – YSI DIQ/S 182 XT-4 System 182 XT-4 User Manual

Page 32

Advertising
background image

Installation

System 182 XT-4

3 - 18

ba76026e01

01/2012

3.6.4

Distributed mounting

General information

For the locally separated connection between DIQ/S 182 XT-4 and
MIQ modules and between MIQ modules, the following cables can be
used:

 SNCIQ cable

 SNCIQ/UG earth cable - suitable for underground laying in

accordance with VDE 01816, Part 2 and DIN/VDE 0891, Part 6.

The cables are delivered as piece goods (please specify length when
ordering!).

SENSORNET

connections

The DIQ/S 182 XT-4 has a SENSORNET connection on the front
terminal strip and two others on the rear terminal strip. The terminal
strips are shown in section 3.14. In addition, free SENSORNET
connections can all be used in the system, e.g. the two SENSORNET
connections of an additional power supply module. All SENSORNET
connections are of equal value in the system.

Caution

The IQ S

ENSOR

N

ET

cable may only be connected to the

SENSORNET connections. No wires of the cable may be
connected with an external electrical potential. Otherwise,
malfunctions could occur.

General installation

instructions

Pay attention to the following points when connecting components via
IQ S

ENSOR

N

ET

lines:

 The sum of all IQ S

ENSOR

N

ET

line lengths (SNCIQ, SNCIQ/UG and

SACIQ) in the system may be up to a maximum of 250.

 IQ S

ENSOR

N

ET

lines must always be installed separately at a

minimum distance of 20 cm from any other lines carrying a voltage
greater than 60 V.

 The terminator switch on the terminal strip of all MIQ modules in the

System 182 must be set to "Off".