Criteria for a stable

measured value

The Stability control function checks whether the measured values are stable
within the monitored time interval.

The minimum duration until a measured value is assessed as stable is the
monitored time interval. The actual duration is mostly longer.

6.2

ORP calibration

You can terminate the Stability control function and the HOLD func-
tion with <AR> or <M> at any time.

2.

Using <MENU/ENTER>, activate the Stability control function manu-
ally.
The [AR] status indicator appears while the measured value is
assessed as not stable. A progress bar is displayed and the display of
the measured parameter flashes.
The [HOLD][AR] status indicator appears as soon as a stable mea-
sured value is recognized. The progress bar disappears and the display
of the measured parameter stops flashing.
The current measurement data is output to the interface. Measurement
data meeting the stability control criterion is marked by AR.

You can prematurely terminate the Stability control function manu-
ally with <MENU/ENTER> at any time. If the Stability control func-
tion is prematurely terminated, the current measurement data are
output to the interface without the AutoRead info.

3.

Using <MENU/ENTER>, start a further measurement with stability con-
trol.
or
Release the frozen measured value again with <AR> or <M>.
The [AR] status display disappears. The display switches back to the
previous indication.

Measured parame-
ter

Time interval

Stability in the time inter-
val

ORP

15 seconds

: Better than 0.3 mV

Temperature

15 seconds

: Better than 0.5 °C

ORP electrodes are not calibrated. You can, however, check ORP
electrodes by measuring the ORP of a test solution and comparing
the value with the nominal value.