Calibration

evaluation

After calibrating, the meter automatically evaluates the calibration. The zero
point and slope are evaluated separately. The worse evaluation of both is taken
into account. The evaluation appears on the display and in the calibration
record.

Calibration record

(USB output)

Display

Calibration
record

Zero point
[mV]

Slope [mV/pH]

+++

-15 ... +15

-60.5 ... -58.0

++

-20 ... <-15
or
>+15 ... +20

>-58.0 ... -57.0

+

-25 ... <-20
or
>+20 ... +25

-61.0 ... <-60.5
or
>-57.0 ... -56.0

-

-30 ... <-25
or
->+25 ... +30

-62.0 ... <-61.0
or
>-56.0 ... -50.0

Clean the IDS sensor according to the sensor operating manual

Error

Error

<-30
or
>+30

<-62.0
or
>-50.0

Error elimination (see section 14 W

HAT

TO

DO

IF

..., page 88)

For pH IDS sensors you can optionally enable a more finely graded
calibration evaluation (QSC) (see section 5.2.8 QSC

FUNCTION

(

SENSOR

QUALITY

CONTROL

), page 40).

CALIBRATION pH
03.04.2013 07:43:33

4110
Ser. no. B092500013
YSI
Buffer 1

4.00

Buffer 2

7.00

Buffer 3

10.00

Voltage 1

184.0 mV 24.0 °C

Voltage 2

3.0 mV 24.0 °C

Voltage

3

-177.0 mV 24.0 °C

Slope

-60.2 mV/pH

Asymmetry

4.0 mV

Sensor

+++