YSI MultiLab 4010P-1 User Manual

Page 31

Advertising
background image

MultiLab 4010-1

pH value

ba76141e03

03/2014

31

Continuing with two-

point calibration

5.

Immerse the sensor in the first buffer solution.

6.

Start the measurement with <MENU/ENTER>.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.

7.

Wait for the end of the measurement with stability control or accept the
calibration value with <MENU/ENTER>.
The calibration display for the next buffer appears (voltage display).

8.

If necessary, finish the calibration procedure as a single-point calibra-
tion with <M>.
The calibration record is displayed.

For single-point calibration, the instrument uses the Nernst slope
(-59.2 mV/pH at 25 °C) and determines the zero point of the IDS
pH sensor.

pH

-180.0

24.8

°C

AutoCal YSI

10.00

01.07.2011

08:00

AR

Buffer

mV

9.

Thoroughly rinse the sensor with deionized water.

10. Immerse the pH sensor in buffer solution 2.

11. When measuring without temperature sensor:

Measure the temperature of the buffer manually and enter it with
<><>.

12. Start the measurement with <MENU/ENTER>.

The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.