YSI MultiLab 4010P-1 User Manual

Page 50

Advertising
background image

Dissolved oxygen

MultiLab 4010-1

50

ba76141e03

03/2014

5.

Start the measurement with <MENU/ENTER>.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter flashes.

6.

Wait for the AutoRead measurement to be completed (audio signal).
The display for the second calibration point is displayed.

7.

Put the IDS D.O. sensor into the calibration medium.

When calibrating via comparison measurement (Comp):
Put the IDS D.O. sensor into the reference solution..

When calibrating in water vapor-saturated air (Cal):
Put the IDS D.O. sensor into the calibration vessel with water vapor-
saturated air.

As the calibration vessel use a BOD bottle that contains a
small amount of clean water (approx. 40 ml).

The sensor must not be immersed in the water.
Leave the sensor in the calibration bottle long enough (5
to 10 minutes) so that the air is water vapor-saturated and
the sensor can adapt to the ambient temperature.

8.

Start the measurement with <MENU/ENTER>.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter flashes.

O

2

0%

0.09

mg/l

Cal

03.04.2013

08:00

760 mmHg

25.1 °C

O

2

DO Sat

0.00

mg/l

Cal

03.04.2013

08:00

760 mmHg

25.1 °C