MultiLab 4010-1

pH value

ba76141e03

03/2014

35

Continuing with

three- to five-point

calibration

14. Wait for the measurement with stability control to be completed or ter-

minate the stability control and take over the calibration value with
<MENU/ENTER>.
The pH value of the buffer solution is displayed.

15. Set the nominal buffer value for the measured temperature with

<><>.

16. Accept the calibration value with <MENU/ENTER>.

The calibration display for the next buffer appears (voltage display).

17. If necessary, finish the calibration procedure as a two-point calibration

with <M>.
The calibration record is displayed.

pH

4.035

24.8

°C

ConCal

2

01.07.2011

08:00

Buffer


18. Thoroughly rinse the sensor with deionized water.

19. Immerse the sensor in the next buffer solution.

20. Start the measurement with <MENU/ENTER>.

The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.

21. Wait for the measurement with stability control to be completed or ter-

minate the stability control and take over the calibration value with
<MENU/ENTER>.
The pH value of the buffer solution is displayed.

pH

9.958

24.8

°C

ConCal

3

01.07.2011

08:00

Buffer