YSI MultiLab 4010P-1 User Manual

Page 58

Advertising
background image

Conductivity

MultiLab 4010-1

58

ba76141e03

03/2014

Calibration

evaluation

After calibration, the meter automatically evaluates the current status of the cal-
ibration. The evaluation appears on the display and in the calibration record.

Calibration record

(USB output)

Calibration /
Calibration data
storage
/ Output to
RS232/USB

-

Outputs the calibration data storage
to the interface

Menu item

Setting/
function

Description

Display

Calibration record

Cell constant [cm

-1

]

+++

Within the range
0.450 ... 0.500 cm

-1

Error

Error

Outside the range
0.450 ... 0.500 cm

-1

Error elimination (see section 14
W

HAT

TO

DO

IF

..., page 88)

CALIBRATION Cond
03.04.2013 07:43:33

4310
Ser. no. 09250033
Cell constant 0.476 1/cm 25.0 °C
Sensor

+++