Settings

MultiLab 4010-1

64

ba76141e03

03/2014

Calibration /
Calibration data
storage
/ Number
of cal. points

1
2

1-point calibration
or
2-point calibration

Calibration /Cali-
bration interval

1 ... 180 ...
999 d

Calibration interval for the IDS D.O.
sensor (in days).
The meter reminds you to calibrate
regularly by the flashing sensor sym-
bol in the measuring screen.

Calibration /Com-
parison meas.

on
off

Enables to adjust the measured value
with the aid of a comparison measure-
ment, e.g. Winkler titration.
For details, see section 7.2 C

ALIBRA

-

TION

, page 47.

Calibration /
Sensor cap coef-
ficients

K1
...
K5
KC

Enter the cap coefficients here after
exchanging the sensor cap.
For details, see section 9.3.2 E

NTER

S

ENSOR

CAP

COEFFICIENTS

, page 65

Salinity/Sal cor-
rection

on
off

Manual salt content correction for con-
centration measurements.

Salinity/Salinity

0.0 ... 70.0

Salinity or salinity equivalent for the
salt content correction.
This menu item is only available if the
manual salt content correction is
switched on.

Resolution

0.1
1

Set a high or low resolution.
The setting of the resolution is sepa-
rately stored for each measured
parameter.

DO % Saturation
local

on
off

DO % Saturation local is a procedure
that takes the local air pressure into
account for each saturation measure-
ment.
For details, see section 9.3.3 DO %
S

ATURATION

LOCAL

, page 65.

Reset

-

Resets all sensor settings to the deliv-
ery condition (see section 9.6.1
R

ESETTING

THE

MEASUREMENT

SET

-

TINGS

, page 70).