Dissolved oxygen

MultiLab 4010-1

52

ba76141e03

03/2014

Calibration

evaluation

After the calibration, the meter automatically evaluates the current status of the
calibration. The evaluation appears on the display and in the calibration record.

ProOBOD

calibration

evaluation

Calibration record

(USB output)

Menu item

Setting/
function

Description

Calibration /
Calibration data stor-
age
/Display

-

Displays the calibration records.

Further options:

Scroll through the calibration
records with <><>.

Output the displayed calibra-
tion record to the interface with
<F2>/[USB output].

Output all calibration records
to the interface with
<F2__>[USB output].

Quit the display with <F1>/
[Back] or <MENU/ENTER>.

Switch directly to the mea-
sured value display with <M>.

Calibration /
Calibration data stor-
age
/ Output to
RS232/USB

-

Outputs the calibration data stor-
age to the interface

Display

Calibration record

Relative slope

+++

S = 0.94 ... 1.06

++

S = 0.92 ... 0.94 or
S = 1.06 ... 1.08

+

S = 0.90 ... 0.92 or
S = 1.08 ... 1.10

Error

Error

S < 0.90 or
S > 1.10

Error elimination (see section 14 W

HAT

TO

DO

IF

...,

page 88)

CALIBRATION Ox
03.04.2013 07:43:33

ProOBOD
Ser. no. 12B100016

ProOBOD Cap

12B100015

Relative slope

0.98

Sensor

+++

Advertising