YSI MultiLab 4010P-1 User Manual
Page 71

MultiLab 4010-1
Settings
ba76141e03
03/2014
71
ORP
The following settings for ORP measurements are reset to the default settings
with the Reset function:
The sensor settings are reset under the Reset menu item in the menu for cali-
bration and measurement settings. To open the settings, display the required
measured parameter in the measured value display and press the <MENU/
ENTER> key.
Dissolved oxygen
The following settings for D.O. measurements are reset to the default settings
with the Reset function:
The sensor settings are reset under the Reset menu item in the menu for cali-
bration and measurement settings. To open the settings, display the required
measured parameter in the measured value display and press the <MENU/
ENTER> key.
Conductivity
The following settings for conductivity measurements are reset to the default
settings with the Reset function:
Setting
Default settings
mV resolution
0.1
Man. temperature
25 °C
Setting
Default settings
Calibration interval
180 d
Measured parameter
D.O. concentration (mg/l)
Salinity (value)
0.0
Salinity (function)
off
Resolution
0.1
DO % Saturation local
off
Setting
Default settings
Calibration interval
150 d
Measured parameter
ϰ
Cell constant (c)
Depending on the connected mea-
suring cell:
0.475 cm
-1
(calibrated)
0.475 cm
-1
(set)
0.100 cm
-1
Temperature compensation
nLF
Reference temperature
25 °C
Temperature coefficient (TC) of the
linear temperature compensation
2.000 %/K
Multiplier for TDS
1.00