What to do if...

MultiLab 4010-1

90

ba76141e03

03/2014

Error message,

Error

14.3 Conductivity

Error message,

OFL, UFL

The measured value is outside the measuring range.

Error message,

Error

14.4 General information

Sensor symbol

flashes

Cause

Remedy

– Sensor contaminated

– Clean the sensor

– Measured temperature value

outside the operating conditions
(display of OFL/UFL instead of a
temperature value)

– Keep to the temperature range for

the test sample

– Defective sensor

– Calibrate
– Change sensor cap
– Replace sensor

More information and instructions on cleaning and exchange of
sensors are given in the documentation of your sensor.

Cause

Remedy

– Measured value outside the

measuring range

– Use suitable IDS conductivity sensor

Cause

Remedy

– Sensor contaminated

– Clean the sensor and replace it if

necessary

– Calibration solution not suitable

– Check the calibration solutions

Cause

Remedy

– Calibration interval expired

– Recalibrate the measuring system