YSI MultiLab 4010P-1 User Manual

Page 32

Advertising
background image

pH value

MultiLab 4010-1

32

ba76141e03

03/2014

Continuing with

three- to five-point

calibration

13. Wait for the measurement with stability control to be completed or ter-

minate the stability control and take over the calibration value with
<MENU/ENTER>.
The calibration display for the next buffer appears (voltage display).

14. If necessary, finish the calibration procedure as a two-point calibration

with <M>.
The calibration record is displayed.

pH

0.0

24.8

°C

AutoCal YSI

7.000

01.07.2011

08:00

AR

Buffer

mV

15. Thoroughly rinse the sensor with deionized water.

16. Immerse the sensor in the next buffer solution.

17. Start the measurement with <MENU/ENTER>.

The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.

18. Wait for the measurement with stability control to be completed or ter-

minate the stability control and take over the calibration value with
<MENU/ENTER>.
The calibration display for the next buffer appears (voltage display).

19. If necessary, use <M> to finish the calibration.

The calibration record is displayed.
or
Switch to calibration with the next buffer with <MENU/ENTER>.

pH

180.0

24.8

°C

AutoCal YSI

4.00

01.07.2011

08:00

AR

Buffer

mV