5.2.5 Calibration points

Calibration can be performed using one to five buffer solutions in any order
(single-point to five-point calibration). The meter determines the following
values and calculates the calibration line as follows:

22. Set the nominal buffer value for the measured temperature with

<><>.

23. Accept the calibration value with <MENU/ENTER>.

The calibration display for the next buffer appears (voltage display).

24. If necessary, use <M> to finish the calibration.

The calibration record is displayed.
or
Continue calibrating using the next buffer with <MENU/ENTER>.

After the fifth buffer has been measured the calibration is automat-
ically finished. Then the calibration record is displayed.

The calibration line is determined by linear regression.

Calibration

Determined values

Displayed calibration data

1-point

Asy

Zero point = Asy

Slope = Nernst slope
(-59.2 mV/pH at 25 °C)

2-point

Asy
Slp.

Zero point = Asy

Slope = Slp.

3-point to 5-
point

Asy
Slp.

Zero point = Asy

Slope = Slp.

The calibration line is calculated
by linear regression.

You can display the slope in the units, mV/pH or % (see section
9.1.1 S

ETTINGS

FOR

P

H

MEASUREMENTS

, page 59).