6 calibration data, Calibration data – YSI MultiLab 4010P-1 User Manual

Page 51

Advertising
background image

MultiLab 4010-1

Dissolved oxygen

ba76141e03

03/2014

51

7.2.6 Calibration data

Displaying the

calibration record

The calibration record of the last calibration is to be found under the menu item,
Calibration / Calibration record. To open it in the measured value display, press
the <CAL__> key.

The calibration records of the last 10 calibrations are to be found in the menu,
Calibration/Calibration data storage / Display. To open the Calibration menu,
press the <MENU/ENTER> key in the measured value display.

9.

Wait for the AutoRead measurement to be completed (audio signal).

When calibrating in water vapor-saturated air (Cal):
The calibration record is displayed and output to the interface.

When calibrating via comparison measurement (Comp):
Set the concentration value with <> <> and accept it with
<ENTER>.
The calibration record is displayed and output to the interface.

10.

Switch to the measured value display with <MENU/ENTER>.

The calibration record is automatically transmitted to the interface
after calibrating.

O

2

DO Adjust

7.92

mg/l

Comp

03.04.2013

08:00


760 mmHg

25.1 °C