8.3.4 Calibration data

The calibration data can be displayed and then output to the interface.

Displaying the

calibration record

The calibration record of the last calibration is to be found under the menu item,
Calibration / Calibration record. To open it in the measured value display, press
the <CAL__> key.

The calibration records of the last 10 calibrations are to be found in the menu,
Calibration / Calibration data storage / Display. To open the Calibration menu,
press the <MENU/ENTER> key in the measured value display.

4.

Immerse the conductivity sensor in the control standard solution,
0.01 mol/l KCI.

5.

Start the measurement with <MENU/ENTER>.
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter flashes.

6.

Wait for the end of the measurement with stability control ([HOLD][AR])
status indicator or
take over the calibrated value with <MENU/ENTER>.
The calibration record is displayed and output to the interface.

7.

Switch to the measured value display with <MENU/ENTER>.

The calibration record is automatically transmitted to the interface
after calibrating.

Menu item

Setting/
function

Description

Calibration /
Calibration data
storage
/ Display

-

Displays the calibration records.

Further options:

Scroll through the calibration
records with <><>.

Output the displayed calibration
record to the interface with <F2>/
[USB output].

Output all calibration records to the
interface with <F2__>[USB out-
put]
.

Quit the display with <F1>/[Back]
or <MENU/ENTER>.

Switch directly to the measured
value display with <M>.