MultiLab 4010-1

Settings

ba76141e03

03/2014

65

9.3.2 Enter Sensor cap coefficients

9.3.3 DO % Saturation local

Irrespective of the height or air pressure, the calibration value is set to 100%.
DO % Saturation local is ideal for EU compliance.

When DO % Saturation local is enabled, an L will appear next to D.O.% on the
display.

D.O. mg/L readings are unaffected by the selection of DO % Saturation local.

The values of the coefficients are provided with the sensor cap.

1.

Change the digit of the highlighted position with <><>.

2.

Go to the next position with <><>.

3.

Confirm with<ENTER> when a coefficient is completely entered.