6 reset, 1 resetting the measurement settings, Reset – YSI MultiLab 4010P-1 User Manual

Page 70: Resetting the measurement settings, On (see sec- tion 9.6.1 r, Esetting, Measurement, Settings, Section 9.6.1 r, Tings

Advertising
background image

Settings

MultiLab 4010-1

70

ba76141e03

03/2014

You can also generally switch on the display illumination (see section 9.5.1
S

YSTEM

, page 68).

9.6

Reset

You can reset (initialize) all sensor settings and sensor-independent settings
separately from each other.

9.6.1 Resetting the measurement settings

pH

The following settings for pH measurements are reset to the default settings
with the Reset function:

The sensor settings are reset under the Reset menu item in the menu for cali-
bration and measurement settings. To open the settings, display the required
measured parameter in the measured value display and press the <MENU/
ENTER>
key.

The calibration data are reset to the default settings together with
the measuring parameters. Recalibrate after performing a reset.

Setting

Default settings

Buffer YSI

Calibration interval

7 d

Unit for slope

mV/pH

Measured parameter

pH

Unit for slope

0.001

mV resolution

0.1

Asymmetry

0 mV

Slope

-59.2 mV

Man. temperature

25 °C

Single-point calibration

off