MultiLab 4010-1

88

ba76141e03

03/2014

14 What to do if...

14.1 pH

Error message

OFL, UFL

Error message,

Error

More information and instructions on cleaning and exchange of
sensors are given in the documentation of your sensor.

Cause

Remedy

IDS pH sensor:

– Measured value outside the

measuring range

– Use suitable IDS pH sensor

– Air bubble in front of the junction – Remove air bubble

– Cable broken

– Exchange IDS pH sensor

– Gel electrolyte dried out

– Exchange IDS pH sensor

Cause

Remedy

IDS pH sensor:

– The values determined for zero

point and slope of the pH sensor
are outside the allowed limits.

– Recalibrate

– pH sensor contaminated

– Clean IDS pH sensor

– pH sensor broken

– Exchange IDS pH sensor

Buffer solutions:

– The used buffer solutions do not

agree with the set buffer set

– Set different buffer set
or
– Use different buffer solutions

– Buffer solutions too old

– Use only once.

Note the shelf life

– Buffer solutions depleted

– Change solutions