No stable measured

value

Obviously incorrect

measured values

14.2 Dissolved oxygen

Error message,

OFL, UFL

The measured value is outside the measuring range.

Cause

Remedy

IDS pH sensor:

– pH sensor contaminated

– Clean pH sensor

Test sample:

– pH value not stable

– Measure with air excluded if neces-

sary

– Temperature not stable

– Adjust temperature if necessary

IDS pH sensor + test sample:

– Conductivity too low

– Use suitable IDS pH sensor

– Temperature too high

– Use suitable IDS pH sensor

– Organic liquids

– Use suitable IDS pH sensor

Cause

Remedy

IDS pH sensor:

– pH sensor unsuitable

– Use suitable IDS sensor

– Temperature difference

between buffer and test sample
too great

– Adjust temperature of buffer or sam-

ple solutions

– Measurement procedure not

suitable

– Follow special procedure

More information and instructions on cleaning and exchange of
sensors are given in the documentation of your sensor.

Cause

Remedy

– Measured value outside the

measuring range

– Use a suitable IDS D.O. sensor