MultiLab 4010-1

Overview

ba76141e03

03/2014

9

have quick-lock couplings with which to fix the sensors to the meter.

Sensor data from

IDS sensors

IDS sensors transmit the following sensor data to the meter:

SENSOR ID

– Sensor name
– Sensor series number

Calibration data

Measurement settings

The calibration data are updated in the IDS sensor after each calibration proce-
dure. A message is displayed while the data are being updated in the sensor.

1.3.2 Automatic sensor recognition

The automatic sensor recognition for IDS sensors allows

to operate an IDS sensor with different meters without recalibrating

to assign measurement data to an IDS sensor

– Measurement datasets are always stored and output with the sensor

name and sensor series number.

to assign calibration data to an IDS sensor

– Calibration data and calibration history are always stored and output with

the sensor name and sensor series number.

to hide menus automatically that do not concern this sensor

To be able to use the automatic sensor recognition, a meter that supports the
automatic sensor recognition (e.g. MultiLab 4010-1) and a digital IDS sensor
are required.

In digital IDS sensors, sensor data are stored that clearly identify the sensor.
The sensor data are automatically taken over by the meter.

In the measured value display, you can display the sensor name
and series number of the selected sensor with the [Info] softkey.
You can then display further sensor data stored in the sensor with
the [More] softkey (see section 4.1.5 S

ENSOR

INFO

, page 18).