15 technical data, 1 measuring ranges, resolution, accuracy, 2 general data – YSI MultiLab 4010P-1 User Manual

Page 92: Section 15 t, Echnical, Data, E section 15 t, See section 15.2 g, Eneral, Section 15.2 g

Advertising
background image

Technical data

MultiLab 4010-1

92

ba76141e03

03/2014

15 Technical data

15.1 Measuring ranges, resolution, accuracy

15.2 General data

Measuring ranges,
accuracy

Parameter

Measuring range

Accuracy

Air pressure (abso-
lute)*

225 ... 825 mm Hg

± 3 mm Hg

*

Available only if a D.O. sensor is connected

Further data are given in the documentation of your sensor.

Dimensions

MultiLab 4010-1:

Approx. 230 x 190 x 80 mm
(9.06 x 7.48 x 3.15 inches)

Weight

MultiLab 4010-1:

Approx. 0.8 kg (1.76 pounds)

Mechanical structure

Type of protection

IP 43

Electrical safety

Protective class

III

Test certificates

CE, cETLus

Ambient conditions

Storage

-25 °C ... +65 °C

Operation

+5 °C ... +55 °C
With the power pack connected:
+5 °C ... +40 °C

Admissible relative
humidity

Yearly mean: < 75 %
30 days/year: 95 %
Other days: 85 %

Power supply

Batteries

4 x 1.5 V alkali-manganese batteries, type
AA

Operational life

Approx. 150 h*

Power pack

Kuantech Co. Ltd. KSAC 0900110W1UV-1
Input: 100 ... 240 V ~ / 50 ... 60 Hz / 270 mA
Output: 9 V = / 1.1 A
Connection max. overvoltage category II
Primary plugs contained in the scope of
delivery: Euro, US, UK and Australian.

*

The operational life is shorter fs the display illumination is switched on permanently

Advertising
This manual is related to the following products: