Ph 1 – Xylem inoLab 7310_7310P WASTEWATER pH METER User Manual
Page 26

pH value
pH 7310
26
ba75929e03
12/2011
Continuing with two-
point calibration
5.
Thoroughly rinse the electrode with deionized water.
6.
Immerse the electrode in the first buffer solution.
7.
When measuring without temperature sensor:
Measure the temperature of the buffer and enter it with <><>.
8.
Start the measurement with <ENTER>.
The measured value is checked for stability (stability control). The [AR]
status indicator is displayed. A progress bar is displayed and the display
of the measured parameter flashes.
9.
Wait for the end of the measurement with stability control or accept the
calibration value with <ENTER>.
The calibration display for the next buffer appears (voltage display).
10. If necessary, finish the calibration procedure as a single-point calibration
with <M>.
The calibration record is displayed.
For single-point calibration, the instrument uses the Nernst slope
(-59.2 mV/pH at 25 °C) and determines the zero point of the elec-
trode.
pH
1
-180.0
24.8 °C
AutoCal TEC
01.07.2011
08:00
Buffer
mV
pH
-180.0
24.8 °C
AutoCal TEC
10.011
01.07.2011
08:00
AR
Buffer
mV
11. Thoroughly rinse the electrode with deionized water.
12. Immerse the electrode in the second buffer solution.
13. When measuring without temperature sensor:
Measure the temperature of the buffer and enter it with <><>.