Xylem inoLab 7310_7310P WASTEWATER pH METER User Manual

Page 30

Advertising
background image

pH value

pH 7310

30

ba75929e03

12/2011

Continuing with

three- to five-point

calibration

14. Immerse the electrode in the second buffer solution.

15. When measuring without temperature sensor:

Measure the temperature of the buffer and enter it with <><>.

16. Start the measurement with <ENTER>.

The measured value is checked for stability (stability control). The [AR]
status indicator is displayed. A progress bar is displayed and the display
of the measured parameter flashes.

17. Wait for the measurement with stability control to be completed or termi-

nate the stability control and take over the calibration value with
<ENTER>

.

The calibration display for the setting of the nominal buffer value appears.

18. Set the nominal buffer value for the measured temperature with

<><>

.

19. Accept the set calibration value with <ENTER>.

The calibration display for the next buffer appears (voltage display).

20. Finish the calibration procedure as a two-point calibration with <M>.

The calibration record is displayed.

pH

4.035

24.8 °C

ConCal

2

01.07.2011

08:00

Buffer



21. Thoroughly rinse the electrode with deionized water.

22. Immerse the electrode in the next buffer solution.

23. When measuring without temperature sensor:

Measure the temperature of the buffer and enter it with <><>.

24. Start the measurement with <ENTER>.

The measured value is checked for stability (stability control). The [AR]
status indicator is displayed. A progress bar is displayed and the display
of the measured parameter flashes.

25. Wait for the measurement with stability control to be completed or termi-

nate the stability control and take over the calibration value with
<ENTER>

.

The calibration display for the setting of the nominal buffer value appears.