pH 7310

What to do if...

ba75929e03

12/2011

59

12

What to do if...

12.1

pH/ORP

Error message

OFL, UFL

The measured value is outside the measuring range.

Error message,

Error

More information and instructions on cleaning and exchange of
sensors are given in the documentation of your sensor.

Cause

Remedy

pH electrode:

– Air bubble in front of the junction

– Remove air bubble

– Air in the junction

– Extract air or moisten junction

– Cable broken

– Replace electrode

– Gel electrolyte dried out

– Replace electrode

– Measured value outside the mea-

suring range of the meter

– Use suitable electrode

Cause

Remedy

pH electrode:

– The values determined for zero

point and slope of the electrode
are outside the allowed limits.

– Recalibrate

– Junction contaminated

– Clean junction

– Electrode broken

– Replace electrode

Buffer solutions:

– Wrong buffer solutions

– Change calibration procedure

– Buffer solutions too old

– Use only once.

Note the shelf life

– Buffer solutions depleted

– Change solutions