2 buffer sets for calibration, Buffer sets for calibration – Xylem inoLab 7310_7310P WASTEWATER pH METER User Manual

Page 39

Advertising
background image

pH 7310

Settings

ba75929e03

12/2011

39

7.1.2

Buffer sets for calibration

You can use the buffer sets quoted in the table for an automatic calibration. The
pH values are valid for the specified temperature values. The temperature
dependence of the pH values is taken into consideration during the calibration.

Calibration

/Buffer

TEC
NIST/DIN
ConCal
...

Buffer sets to be used for pH cali-
bration (see section 5.2 C

ALIBRA

-

TION

, page 25).

Calibration

/One

point calibration

Yes
No

Quick calibration with 1 buffer

Calibration

/Cali-

bration interval

1 ... 7 ... 999 d

Calibration interval

for the pH elec-

trode (in days).
The meter reminds you to calibrate
regularly by the flashing sensor
symbol in the measuring window.

Calibration

/Unit for

slope

mV/pH
%

Unit of the slope.
The % display refers to the Nernst
slope of -59.2 mV/pH (100 x deter-
mined slope/Nernst slope).

Man. temperature

-25 ... +25 ...
+130 °C

Entry of the manually determined
temperature. For measurements
without temperature sensor only.

Temperature unit

°C

°F

Temperature unit,

degrees Celsius or degrees Fahr-
enheit.
All temperatures are displayed
with the selected unit.

Resolution pH

0.001
0.01
0.1

Resolution of the pH display

Stability control

On
Off

Switches on or off the automatic
stability control during measure-
ment (see section 5.1.1 M

EASUR

-

ING

THE

P

H

VALUE

, page 22)

Reset

-

Resets all sensor settings to the
delivery condition (see section
7.4.1 R

ESETTING

THE

MEASURE

-

MENT

SETTINGS

, page 44).

Menu item

Possible set-
ting

Description