3 calibration interval, 2 settings for orp measurements, Calibration interval – Xylem inoLab 7310_7310P WASTEWATER pH METER User Manual
Page 42: Settings for orp measurements

Settings
pH 7310
42
ba75929e03
12/2011
7.1.3
Calibration interval
The calibration evaluation is displayed as a sensor symbol.
The sensor symbol flashes after the adjusted calibration interval has expired. It
is still possible to measure.
Setting the calibra-
tion interval
The calibration interval is set to 7 days in the factory.
You can change the interval (1 ... 999 days):
7.2
Settings for ORP measurements
The settings are made in the menu for calibration and measurement settings of
the pH/ORP measurement. To open the settings, display the required parame-
ter in the measured value display and press the <F1>/[menu] or <ENTER> key.
After completing the settings, switch to the measured value display with <M>.
Default settings are printed in bold.
The buffers are selected in the menu, pH / <F1>/[Menu] / Calibra-
tion
/ Buffer (see 7.1.1 C
38).
To ensure the high measuring accuracy of the measuring system,
calibrate after the calibration interval has expired.
1.
Open the menu for measurement settings with [Menu]<F1>.
2.
In the Calibration / Calibration interval menu, set the calibration interval
with <><>.
3.
Confirm the setting with <ENTER>.
4.
Quit the menu with <M>.
Menu item
Possible
setting
Description
Man. temperature
-25 ... +25
... +130 °C
Entry of the manually determined tem-
perature. For measurements without
temperature sensor only.
Temperature unit
°C
°F
Temperature unit,
degrees Celsius or degrees Fahrenheit.
All temperatures are displayed with the
selected unit.