background image

pH 7310

pH value

ba75929e03

12/2011

33

Calibration record

(example)

Display

Calibration
record

Zero point
[mV]

Slope

[mV/pH]

+++

- 15 ... + 15

-60,5 ... -58,0

++

- 20 ... <-15
or
>+15 ... + 20

>-58,0 ... -57,0

+

- 25 ... <- 20
or
>+ 20 ... + 25

-61,0 ... <-60,5
or
>-57,0 ... -56,0

-

- 30 ... <- 25
or
->+25 ... + 30

-62,0 ... <-61,0
or
>-56,0 ... -50,0

Clean the electrode according to the
electrode operating manual

Error

Error

- 30 ... + 30

-62,0 ... -50,0

Error elimination (see section 12 W

HAT

TO

DO

IF

..., page 59)

inoLab pH 7310
Ser. no. 11292113

CALIBRATION pH
30.06.2011 15:55

Ser. no. 10501234
AutoCal TEC
Buffer 1

4.01

Buffer 2

7.00

Buffer 3

10.01

Voltage

1

184.0 mV 24.0 °C

Voltage

2

3.0 mV 24.0 °C

Voltage 3

-177.0 mV 24.0 °C

Slope

-60.2 mV/pH

Asymmetry

4.0 mV

Sensor

+++

etc...