2 data storage, 4 reset, 1 resetting the measurement settings – Xylem inoLab 7310_7310P WASTEWATER pH METER User Manual

Page 44: Data, Reset, Resetting the measurement settings, Section 7.4.1 r, Esetting, Measure, Ment

Advertising
background image

Settings

pH 7310

44

ba75929e03

12/2011

7.3.2

Data storage

This menu contains all functions to display, edit and erase stored measured
values and calibration records (see section 8 D

ATA

STORAGE

, page 46).

7.4

Reset

You can reset (initialize) all sensor settings and sensor-independent settings
separately from each other.

7.4.1

Resetting the measurement settings

The following settings for pH measurements are reset to the default settings
with the Reset function:

System

/ Interface /

Output header

Output of a header for Output
format

: CSV

System

/ Printer

Settings for the integrated prin-
ter of the pH 7310 P.
For details, see section 10
P

RINTER

(

ONLY

P

H 7310P),

page 54

System

/Clock

Date format
Date
Time

Time and date settings (see
section 4.4.6 E

XAMPLE

2

ON

NAVIGATION

: S

ETTING

THE

DATE

AND

TIME

, page 20)

System

/Service

information

Hardware version and software
version of the meter are dis-
played.

System

/Reset

-

Resets the system settings to
the delivery condition (see sec-
tion 7.4.2 R

ESETTING

THE

SYS

-

TEM

SETTINGS

, page 45).

The calibration data are reset to the default settings together with
the measuring parameters. Recalibrate after performing a reset.

Setting

Default settings

Buffer

AutoCalTEC

Cal. interval

7 d

Unit for slope

mV/pH

Menu item

Possible set-
ting

Description