Xylem inoLab 7310_7310P WASTEWATER pH METER User Manual

Page 29

Advertising
background image

pH 7310

pH value

ba75929e03

12/2011

29

Continuing with two-

point calibration

6.

Immerse the electrode in the first buffer solution.

7.

When measuring without temperature sensor:
Measure the temperature of the buffer and enter it with <><>.

8.

Start the measurement with <ENTER>.
The measured value is checked for stability (stability control). The [AR]
status indicator is displayed. A progress bar is displayed and the display
of the measured parameter flashes.

9.

Wait for the end of the measurement with stability control or accept the
calibration value with <ENTER>.

The calibration display for the setting of the nominal buffer value appears.

10. Set the nominal buffer value for the measured temperature with

<><>

.

11. Accept the set calibration value with <ENTER>.

The calibration display for the next buffer appears (voltage display).

12. If necessary, finish the calibration procedure as a single-point calibration

with <M>.
The calibration record is displayed.

For single-point calibration, the instrument uses the Nernst slope

(-59.2 mV/pH at 25 °C) and determines the zero point of the elec-
trode.

pH

0.0

24.8 °C

ConCal

7.000

01.07.2011

08:00

Buffer

mV

AR

pH

7.000

24.8 °C

ConCal

1

01.07.2011

08:00

Buffer



13. Thoroughly rinse the electrode with deionized water.

Advertising