7 settings, 1 measurement settings, 1 changing the settings for ph measurements – Xylem inoLab 7310_7310P WASTEWATER pH METER User Manual

Page 38: Settings, Measurement settings, Changing the settings for ph measurements, Section 7.1.1 c, Hanging, Measurements, Ee section 7.1.1 c

Advertising
background image

Settings

pH 7310

38

ba75929e03

12/2011

7

Settings

7.1

Measurement settings

7.1.1

Changing the settings for pH measurements

The settings are made in the menu for calibration and measurement settings of
the pH/ORP measurement. To open the settings, display the required parame-
ter in the measured value display and press the <F1>/[menu] or <ENTER> key.
After completing the settings, switch to the measured value display with <M>.
Default settings are printed in bold.

Menu item

Possible set-
ting

Description

Calibration

/Cali-

bration record

-

Displays the calibration record of
the last calibration.

Calibration

/

Calibration data
storage

/Display

-

Displays the calibration record.

Further options:

Scroll through the calibration
records with <><>.

Output the displayed calibration
record to the interface with
<F2>

/[USB output].

Output all calibration records to
the interface with <F2__>[USB
output]

.

Quit the display with <F1>/
[Back]

or <ENTER>.

Switch directly to the measured
value display with <M>.

Calibration

/ Cali-

bration data stor-
age

/ Output to USB

-

Outputs the calibration records to
the interface.

Calibration

/ Serial

number (probe)

Entry of the serial number of the
connected sensor.
The serial number is output in the
calibration record.

Change the content of the high-
lighted position with <><>.

Go to the next position with
<F2>

/[].

When the serial number has
been completely entered, con-
firm with <ENTER>.