3 ph value / orp voltage, 1 general information, Ph value / orp voltage 4.3.1 – Xylem MULTI 1970I PORTABLE MULTIPARAMETER METERS User Manual
Page 19: General information

Multi 1970i
Operation
17
4.3
pH value / ORP voltage
4.3.1
General information
Preparatory activities
Perform the following preparatory activities when you want to measure:
Note
Incorrect calibration of pH electrodes leads to incorrect measured val-
ues. Calibrate regularly before measuring. You can only connect elec-
trodes of the NTC30 type or without temperature sensor.
Warning
When connecting an earthed PC/printer, measurements cannot be per-
formed in earthed media as incorrect values would result. The RS232
interface is not galvanically isolated.
1
Connect the pH depth armature or the pH electrode to the mea-
suring instrument.
If necessary, press the <M> key repeatedly until the pH (pH
measurement) or U (measurement of the ORP voltage) display
appears.
2
Adjust the temperature of the buffer solutions or test solutions,
or measure the current temperature, if you measure without a
temperature sensor.
3
Calibrate or check the measuring instrument with the elec-
trode.
4
Using <
▲ > <▼>, toggle between the pH or mV measuring
modes.