Multi 1970i

Operation

29

4.4.2

Measuring the D. O. concentration

When measuring the concentration of test samples with a salt content
of more than 1 g/l, a salinity correction is required.

Note
How to enter the current salt content is described in
section 4.4.6 E

NTERING

THE

SALT

CONTENT

(

SALINITY

) on page 35.

Switching the salt content correction on or off, see below.

To measure the D. O. concentration with and without salt content cor-
rection, proceed as follows:

Switching on/off the salt

content correction

Proceed as follows to switch on the salt content correction:

1

Perform the preparatory activities according to section 4.4.1.

2

Immerse the D. O. sensor in the test sample.

3

Press the <

> <> key repeatedly until the D. O. concentra-

tion in mg/l appears on the display.

Sal

S

pH

AR

1

LoBat

nLF

Lin

Oxi

Tref25

Tref20

Cal

TEC

Auto

Store

8

Time

Year

Ident

No.

Day.Month

Baud

Sal

cm

1/

K

/

%

RCL

REL2

pH

/

mV

m

cm

S/

%

7

1 7 6

9 2

°

C

TP

REL1

O

ARng

mg/ l

1

Perform the preparatory activities according to section 4.4.1.

2

Immerse the D. O. sensor in the test sample.

3

While pressing the <RUN/ENTER> key, switch on the salt con-
tent correction with <

> . The SAL display indicator appears

on the display. The specified salt content is taken into consid-
eration during the measurement.