Multi 1970i

What to do if...

69

6.2

Oxi system messages

Error message

0FL

0FL

0FL

0FL

Error message

E3

E3

E3

E3

Error message

E7

E7

E7

E7

AR flashes

continuously

Cause

Remedy

Display range exceeded

Oxygen sensor

– Not connected

– Connect the sensor

– Cable broken

– Replace sensor

– Depleted

– Replace sensor

– Short-circuit between gold and

lead electrode

– Clean sensor and replace it if

necessary

Cause

Remedy

Invalid calibration

Oxygen sensor

– Electrolyte solution depleted

– Regenerate sensor

– Membrane contaminated

– Clean membrane

– Electrode system poisoned

– Regenerate sensor

– Obsolete

– Replace sensor

– Broken

– Replace sensor

Cause

Remedy

Membrane damaged

– Membrane damaged

– Regenerate sensor

– Membrane head screwed on

not tight enough

– Screw membrane head tight

Cause

Remedy

No stable measured value

– Membrane contaminated

– Clean membrane