4.4.6

Entering the salt content (salinity)

A salt content correction is required in the oxygen concentration mea-
surement of samples with a salt content of more than 1 g/l. To do this,
you have to enter the salinity equivalent (the measured salinity) of the
test sample (range 0.0 - 70.0) and to switch on the salinity correction.

Note
With the Multi 1970i, you can measure the salinity. How to proceed is
described in section 4.5.3 M

EASURING

THE

SALINITY

on page 38.

Entering the

salt content

Note
How to switch on the salt content correction is described on page 29.

1

Determine the salinity of the test sample (any method, see also
section 4.5.3 M

EASURING

THE

SALINITY

on page 38).

2

Press the <CAL> key repeatedly until Sal appears on the dis-
play.

3

Enter the salt content with <

> <>.

4

Switch to the measuring mode with <M>.