6 calibration intervals (int3, int4, int5), Calibration intervals (int 3, int 4, int 5), 4n t – Xylem MULTI 1970I PORTABLE MULTIPARAMETER METERS User Manual

Page 45

Advertising
background image

Multi 1970i

Operation

43

4.6

Calibration intervals (Int 3, Int 4, Int 5)

For each measured variable, a time interval is stored. When it has ex-
pired, you will be reminded to calibrate. After a calibration interval has
expired, the sensor symbol of the relevant measured variable flashes.
It is still possible to measure. By calibrating the relevant sensor, the
function is reset and the interval starts anew.

The following calibration intervals are set in the factory:

Setting the calibration

interval

You can change each of these intervals (1 ... 999 days):

Measured
parameter

Designation

Default setting

pH/ORP

Int 3

7 days

Dissolved
oxygen

Int 4

14 days

Conductivity

Int 5

180 days

1

Switch off the measuring instrument.

2

Press the <M> key and hold it down.

3

Press the <ON/OFF> key.
The display test appears briefly on the display. After this, the
measuring instrument automatically switches over to configu-
ration.

4

Press the <RUN/ENTER> key repeatedly, until Int 3 ... 5 to-
gether with the required measured variable (pH, O

2

or

) ap-

pears. Example:

5

Set the required interval until the next calibration with <

>

<

>.

6

Confirm with <RUN/ENTER>.

7

Switch to the measuring mode with <M>.

AR

cm

1/

4

Tref25

TP

ARng

nLF

°

C

cm

S/

Sal

RCL

%

S

%

m

mg/ l

Oxi

Year

Ident

No.

Day.Month

Tref20

LoBat

Lin

Sal

pH

TEC

Auto

Store

Baud

K

/

REL2

pH

/

mV

1 4

4

n t

REL1

d

I

O

Time

Cal

Advertising