Calibration evaluation

After calibrating, the measuring instrument automatically evaluates the
calibration. The asymmetry and slope are evaluated separately. The
worst evaluation appears on the display.

Preparatory activities

Display

Asymmetry
[mV]

Slope
[mV/pH]

-15 ... +15

-60.5 ... -58

-20 ... +20

-58 ... -57

-25 ... +25

-61 ... -60.5
or
-57 ... -56

Clean the electrode according
to the electrode operating
manual

-30 ... +30

-62 ... -61
or
-56 ... -50

E3

Perform error elimination ac-
cording to chapter 6 W

HAT

TO

DO

IF

...

< -30 or
> 30

< -62 or
> -50

1

Connect the pH electrode to the measuring instrument.
If necessary, press the <M> key repeatedly until the status dis-
play pH (pH measurement) or U (measurement of the ORP
voltage) appears.

2

Keep the buffer solutions ready.

3

Adjust the temperature of the solution and measure the current
temperature if the measurement is made without the use of a
temperature sensor (the TP display indicator is missing from
the display).