Xylem MULTI 1970I PORTABLE MULTIPARAMETER METERS User Manual

Page 26

Advertising
background image

Operation

Multi 1970i

24

AutoCal TEC

For this procedure, use any two WTW technical buffer solutions (pH
values at 25 °C: 2.00 / 4.01 / 7.00 / 10.01).

Note
The calibration for pH 10.01 is optimized for the WTW technical buffer
solution TEP 10 Trace or TPL 10 Trace. Other buffer solutions can lead
to an erroneous calibration. The correct buffer solutions are given in the
WTW catalog or in the Internet.

Note
The buffer solutions are automatically recognized by the measuring in-
strument. Depending on the instrument setting (see
section 4.9 C

ONFIGURATION

), the instrument displays the relevant buff-

er nominal value or the current electrode voltage in mV.

Note
Skip the steps 2 and 7 if you use a pH electrode with temperature sen-
sor or the temperature sensor of a conductivity measuring cell or a D.
O. sensor.

Starting the calibration

1

Press the <CAL> key. The Ct1 display and the function display
AutoCal TEC appears. The sensor symbol displays the evalu-
ation of the last calibration (or no sensor symbol in the delivery
state or after the measurement parameters have been reset).

2

If required, enter the temperature of the first buffer solution with
<

> <> while keeping the <RUN/ENTER> key depressed.

3

Immerse the pH electrode in the first buffer solution.

4

Press the <RUN/ENTER> key.
The AR display indicator flashes.
The electrode voltage (mV) or the buffer nominal value ap-
pears on the display. Example:

Store

AR

1

LoBat

nLF

Lin

Oxi

Tref25

Tref20

ARng

8

Time

Year

Ident

No.

Day.Month

Baud

Sal

cm

1/

K

/

%

RCL

REL2

pH

/

mV

m

cm

S/

%

mg/ l

Sal

O

S

pH

C

2 4 8

T 1

°

C

TP

REL1

Cal

TEC

Auto