4 general errors, General errors – Xylem MULTI 1970I PORTABLE MULTIPARAMETER METERS User Manual

Page 73

Advertising
background image

Multi 1970i

What to do if...

71

6.4

General errors

Display LoBat

Instrument does not

react

to keystroke

Display

to

to

to

to

Sensor symbol flashes

Message

StoFull

StoFull

StoFull

StoFull

You want to know which

software version is in

the instrument

Cause

Remedy

– Battery almost empty

– Charge the battery (see

section 3.2)

Cause

Remedy

– Operating condition undefined

or EMC load unallowed

– Processor reset:

Switch the instrument on
while pressing the <RCL> key

Cause

Remedy

– Time-out of the interface

– Check the instrument

connected

Cause

Remedy

– Calibration interval expired

– Recalibrate the measuring

system

Cause

Remedy

– All memory locations are full

– Output data storage and clear

data storage

Cause

Remedy

– E. g., a question by the WTW

service department

– Simultaneously press the

<CAL> and <ON/OFF> keys
and release them again. The
software version is displayed.