Outputting the calibra-

tion protocol on the

interface

Sample printout:

1

Press the <RCL> key repeatedly until CAL SEr appears on the
display.

2

Press the <RUN/ENTER> key.
The calibration protocol for all measured variables is transmit-
ted to the interface. After the data transmission, the instrument
automatically switches to the last active measurement mode.

d

Sal

mg/ l

ARng

O

TP

Oxi

Cal

AR

°

C

S

Year

Ident

No.

Day.Month

Tref25

Tref20

LoBat

Lin

nLF

Sal

pH

C

TEC

Auto

Store

Time

Baud

cm

1/

K

/

%

REL2

pH

/

mV

m

cm

S/

%

C

S E R

A L

REL1

RCL

CALIBRATION PROTOCOL
02.03.02 14:19
Device No.: 12345678

Calibration pH
Cal time: 01.03.02 / 15:20
Cal interval: 7d
AutoCal TEC Tauto
Buffer 1 2.00
Buffer 2 4.01
Buffer 3 7.00 *
Buffer 4 10.01
C1 174.1 mV 25.0°C
C2 -133.3 mV 25.0°C
S1 -59.4 mV/pH
ASY1 - 4 mV
Probe: +++

CALIBRATION 02
Cal time: 02.03.02 / 14:19
Cal interval: 14d
OxiCal Tauto AR
Relative Slope: 0,88
Probe: +++

Calibration Conductivity
Cal time: 14.01.02 / 11:37
Cal interval: 180d
Cal Std.: 0.01 mol/l KCL
40.0 °C
Conduct./Tref25: 1413µS/cm
Cell Const : 0.478 1/cm
Probe : +++