Multi 1970i

Technical data

75

7.2.2

Dissolved oxygen

Measuring ranges

and resolutions

Variable

Measuring range

Resolution

Concentration [mg/l]

0 ... 19.99
0 ... 90.0

0.01
0.1

Saturation [%]

0 ... 199.9
0 ... 600

0.1
1

T [°C]

0 ... 50.0

0.1

Accuracies

Variable

Accuracy (± 1 Digit)

Concentration [mg/l]

± 0.5 % of measured value
at ambient temperature of 5 °C ... 30 °C

Saturation [%]

± 0.5 % of measured value
when measuring in the range of ± 10 K
around the calibration temperature

T [°C]

± 0.1

Correction

functions

Temperature com-
pensation

Accuracy better than 2 % at 0 ... 40 °C

Salinity correction

0 ... 70.0 SAL

Air pressure
correction

Automatic through integrated pressure sen-
sor in the range of 500 ... 1100 mbar