C a l – Xylem MULTI 1970I PORTABLE MULTIPARAMETER METERS User Manual

Page 43

Advertising
background image

Multi 1970i

Operation

41

Calibration evaluation

After the calibration, the measuring instrument automatically evaluates
the current status of the calibration. The evaluation appears on the dis-
play.

Determining the

cell constant

You can determine the cell constant (conductivity measuring cell
TetraCon 325) as follows:

Display

Cell constant [cm

-1

]

0.450 ... 0.500 cm

-1

E3

Perform error elimination ac-
cording to chapter 6 W

HAT

TO

DO

IF

...

Outside the range
0.450 ... 0.500 cm

-1

1

Press the <CAL> key repeatedly until the status display

CAL appears.

2

Press the <RUN/ENTER> key. The CAL display appears, as
well as
– the current, calibrated cell constant (with sensor symbol on

the display) or

– the fixed cell constant 0.475 1/cm (without sensor symbol

on the display). In this case, the measurement parameters
are initialized. See “Reset” on page 62..

3

Immerse the conductivity measuring cell in the control standard
solution, 0.01 mol/l KCI.

TP

TP

AR

TEC

Auto

°

C

pH

°

C

/ pH

m

cm

S/

%

mg/ l

mV

pH

AR

ARng

TEC

Auto

mV

S

/

1

Sal

O

LoBat

nLF

Lin

Oxi

Tref20

ARng

Time

Year

Ident

No.

Day.Month

Baud

Sal

K

/

%

RCL

REL2

pH

m

cm

S/

%

mg/ l

4 7 5

REL1

1

Sal

O

S

LoBat

nLF

Lin

Oxi

Tref20

Time

Year

Ident

No.

Day.Month

Baud

Sal

K

/

%

RCL

REL2

L

REL1

C A L

Tref25

Cal

0 4 7 2

A

C

cm

1/

0

Tref25

Cal

cm

1/

Advertising